BOSTON SCIENTIFIC CORPORATION LEVEEN SUPERSLIM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 36705
Medical Device Problem Codes Device Displays Incorrect Message (2591); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: initial reporter address 1: (b)(6).
 
Event or Problem Description
It was reported that the needle kinked.A 2.0/17/15 leveen superslim was selected for use to treat liver cancer.During the procedure, the device frequently reported an alert and stopped working.It was found that the needle was kinked after it was withdrawn.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
E1: initial reporter address 1: (b)(6).Device evaluated by mfr: the electrode device was returned, and no damages were observed at the handle.Visual inspection revealed that the needles were bent.Functional inspection was performed by extending and retracting the handle, and the needle/tines were activated without any problems.An electrical evaluation was performed by measuring the continuity, the electrode was able to conduct current indicating proper connectivity.The resistance test was within specification.
 
Event or Problem Description
It was reported that the needle kinked.A 2.0/17/15 leveen superslim was selected for use to treat liver cancer.During the procedure, the device frequently reported an alert and stopped working.It was found that the needle was kinked after it was withdrawn.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEEN SUPERSLIM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18335272
Report Number2124215-2023-71204
Device Sequence Number12459356
Product Code GEI
UDI-Device Identifier08714729441694
UDI-Public08714729441694
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2025
Device Model Number36705
Device Catalogue Number36705
Device Lot Number0028700979
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2023
Supplement Date Received by Manufacturer01/15/2024
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date02/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient Weight69 KG
-
-