| Catalog Number |
366593 |
| Medical Device Problem Codes |
Product Quality Problem (1506); Difficult or Delayed Activation (2577); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that bd microtainer® contact-activated lancet, needle tip does not rebound, resulting in needlestick injury post-use.This has occurred three times.No additional information reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Oem manufacturing site: the manufacturing location for this product is htl.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.E.1.Initial reporter phone #: (b)(6).H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd microtainer® contact-activated lancet, needle tip does not rebound, resulting in needlestick injury post-use.This has occurred three times.No additional information reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: "material #: 366593.Lot/batch #: c4n51l3.Bd had not received samples or photos for investigation.Therefore, retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to retraction failure as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode retraction failure.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.
|
| |
|