• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE Back to Search Results
Catalog Number 366593
Medical Device Problem Codes Product Quality Problem (1506); Difficult or Delayed Activation (2577); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that bd microtainer® contact-activated lancet, needle tip does not rebound, resulting in needlestick injury post-use.This has occurred three times.No additional information reported.
 
Additional Manufacturer Narrative
Oem manufacturing site: the manufacturing location for this product is htl.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.E.1.Initial reporter phone #: (b)(6).H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd microtainer® contact-activated lancet, needle tip does not rebound, resulting in needlestick injury post-use.This has occurred three times.No additional information reported.
 
Additional Manufacturer Narrative
H.6.Investigation summary: "material #: 366593.Lot/batch #: c4n51l3.Bd had not received samples or photos for investigation.Therefore, retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to retraction failure as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode retraction failure.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® CONTACT-ACTIVATED LANCET
Common Device Name
MANUAL SURGICAL INSTRUMENT FOR GENERAL USE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18335678
Report Number2243072-2023-02178
Device Sequence Number8415801
Product Code FMK
UDI-Device Identifier50382903665933
UDI-Public(01)50382903665933
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K223243
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number366593
Device Lot NumberC4N51L3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2023
Supplement Date Received by Manufacturer01/03/2024
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date01/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-