ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY Back to Search Results
Catalog Number 08445699190
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the c503 is (b)(6).The serial number of the c303 analyzer is (b)(6).A new lot calibration was performed on one of the instruments and the control recovery improved.The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they have been having ongoing issues with patient samples tested with tina-quant hemoglobin a1cdx gen.3 on a cobas pro c 503 analytical unit and a cobas pure c 303 analytical unit.The c503 instrument generally measured hba1c values that were around 1 % higher than other hospitals, particularly for higher values.The customer tested one patient sample for hba1c on the c503 and c303 instruments.The c503 hba1c result was 7.74 % and the c303 result was 7.08 %.The customer stated that the actual value based on clinical symptoms should be about 7.3 %.
 
Additional Manufacturer Narrative
The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TINA-QUANT HEMOGLOBIN A1CDX GEN. 3
Common Device Name
GLYCOSYLATED HEMOGLOBIN ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18336239
Report Number1823260-2023-04044
Device Sequence Number6739803
Product Code LCP
UDI-Device Identifier07613336157824
UDI-Public07613336157824
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K072714
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08445699190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/05/2023
Supplement Date Received by Manufacturer12/25/2023
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date01/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-