ROCHE DIAGNOSTICS ELECSYS TSH; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Catalog Number 08443432190
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The analyzer serial number is (b)(6).The patient sample was requested for investigation.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable elecsys tsh assay results for 1 patient sample on a cobas e 801 analytical unit compared to an abbott architect analyzer.Refer to the attachment to the medwatch for all patient data.
 
Additional Manufacturer Narrative
The qc recovery data provided was acceptable.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18340406
Report Number1823260-2023-04056
Device Sequence Number17827595
Product Code JLW
UDI-Device Identifier07613336157633
UDI-Public07613336157633
Combination Product (Y/N)Y
Initial Reporter CountryTH
PMA/510(K) Number
K961491
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08443432190
Device Lot Number70629800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/21/2023
Supplement Date Received by Manufacturer02/08/2024
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date02/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARBOPLATIN.; ETHAMBUTOL.; ISONIAZID.; PYRAZINAMIDE.; RIFAMPIN.
Patient SexUnknown
-
-