| Catalog Number |
08443432190 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The patient sample was requested for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys tsh assay results for 1 patient sample on a cobas e 801 analytical unit compared to an abbott architect analyzer.Refer to the attachment to the medwatch for all patient data.
|
| |
|
Additional Manufacturer Narrative
|
|
The qc recovery data provided was acceptable.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.
|
| |
|