|
Complaint confirmed.One unpackaged ar-8380sr serial/batch number (b)(6) was received for evaluation.Visual inspection found that the outer tube was bent, and the inside tube was broken off inside the outer tube.The most likely cause(s) for the reported failure can be a user error due to prying/levering the device during insertion, according to dfu-0215 at revision 1.F.Precautions.6.Do not allow the rotating portion of any device to touch any metallic object, such as a cannula or arthroscope, during use.Damage to both devices is likely.Damage to the device can range from a slight distortion or dulling of the blade/burr edge to actual fracture of the tip in vivo.If such contact does occur, inspect the tip.If there are cracks, fractures, or dulling, or if there is any other reason to suspect a blade is damaged, replace it immediately.7.Excessive "side-loading" on the device during use does not improve cutting performance and, in extreme cases, will cause irreversible damage to the device.
|