MAUDE Adverse Event Report: BIOMET UK LTD. OXF TRL BRG W/SLOTS MED 4MM; BMET GENERAL KNEE INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF TRL BRG W/SLOTS MED 4MM; BMET GENERAL KNEE INSTRUMENTS Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2 - foreign: poland.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that a trial knee bearing fractured during a procedure.No piece fell or was left in the patient.;the damaged trial bearing was returned to the distributor warehouse.This event is related to a malfunction that could potentially lead to a sterility issue/serious injury.However, no patient harm or further outcome was reported.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).Visual examination of the returned product identified that the part was returned in 2 pieces.Minor scratches noticed on the part.No other damage noticed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device is used for treatment.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF TRL BRG W/SLOTS MED 4MM
Common Device Name
BMET GENERAL KNEE INSTRUMENTS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18344314
Report Number3002806535-2023-00434
Device Sequence Number17484262
Product Code NRA
UDI-Device Identifier05019279346820
UDI-Public(01)05019279346820(11)230724(10)ZB7318269
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number32-422704
Device Lot NumberZB7318269
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2023
Supplement Date Received by Manufacturer02/29/2024
Initial Report FDA Received Date12/18/2023
Supplement Report FDA Received Date02/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-