• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LNR, +5 LAT 32MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LNR, +5 LAT 32MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 136-32-53
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Cyst(s) (1800); Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377)
Date of Event 11/23/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section h10: (h3) pending evaluation (d10) concomitant device(s): 5145289 180-01-56 - crown cup,cluster-hole gr.56.
 
Event or Problem Description
It was reported that 5.5 years post op this patient presented back to the surgeons office to get left hip prosthesis from 2018 checked.X-ray control showed a slight decentering of the prosthesis head and osteolysis in the acetabulum as signs of inlay wear.This was verified when the inlay was changed on (b)(6) 2023 to a vitd-hardened, specially approved inlay.As part of the replacement operation, in addition to the inlay replacement, the firm socket integrity was determined as well as the curettage and sealing of the cysts in the socket using allogeneic cancellous bone and the replacement of the prosthetic head.
 
Additional Manufacturer Narrative
H3: based on the available information, the patient involved meets the following risk criteria for early prosthesis wear/osteolysis as specified in the hhe: implanted with a lateralized liner.The most likely cause for the revision/monitoring reported due to early prosthesis wear and osteolysis is a combination of the risk factors specified in the hhe.However, this cannot be confirmed from the reported information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Based on the available information, the patient involved in incident meets the following risk criteria for early prosthesis wear and/or osteolysis: implanted with a lateralized liner.The most likely cause for the reported revision due to prosthesis wear and osteolysis is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, +5 LAT 32MM ID, GROUP 3 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
miguel sosa
2320 nw 66 ct
gainesville, FL 32653
3523771140
MDR Report Key18346422
Report Number1038671-2023-03006
Device Sequence Number6212629
Product Code JDI
UDI-Device Identifier10885862024299
UDI-Public10885862024299
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/24/2022
Device Catalogue Number136-32-53
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer02/08/2024
02/06/2025
03/04/2025
Initial Report FDA Received Date12/18/2023
Supplement Report FDA Received Date02/20/2024
02/27/2025
03/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
Patient SexMale
-
-