| Catalog Number |
136-32-53 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377)
|
| Date of Event |
11/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) pending evaluation (d10) concomitant device(s): 5145289 180-01-56 - crown cup,cluster-hole gr.56.
|
| |
|
Event or Problem Description
|
|
It was reported that 5.5 years post op this patient presented back to the surgeons office to get left hip prosthesis from 2018 checked.X-ray control showed a slight decentering of the prosthesis head and osteolysis in the acetabulum as signs of inlay wear.This was verified when the inlay was changed on (b)(6) 2023 to a vitd-hardened, specially approved inlay.As part of the replacement operation, in addition to the inlay replacement, the firm socket integrity was determined as well as the curettage and sealing of the cysts in the socket using allogeneic cancellous bone and the replacement of the prosthetic head.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: based on the available information, the patient involved meets the following risk criteria for early prosthesis wear/osteolysis as specified in the hhe: implanted with a lateralized liner.The most likely cause for the revision/monitoring reported due to early prosthesis wear and osteolysis is a combination of the risk factors specified in the hhe.However, this cannot be confirmed from the reported information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Based on the available information, the patient involved in incident meets the following risk criteria for early prosthesis wear and/or osteolysis: implanted with a lateralized liner.The most likely cause for the reported revision due to prosthesis wear and osteolysis is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|