| Catalog Number |
EVLTPVAK |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Event or Problem Description
|
|
An end user reported an issue with a pvak -- 400 micron fiber procedure kit.Twenty seconds into the initial firing of the fiber, it began popping and sparking, approximately 7 inches from the tip.The fiber was inspected, prior to use, and there was no visible damage noted.The portion of the laser, where this occurred, was not inside the patient at the time of the popping and sparking.The generator was set at 7w and no error messages occurred.The operator experienced a small, superficial burn on their finger; however, they did not require medical intervention as a result of this incident and the procedure was completed with another same device.The patient did not experience any adverse effects or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's reported complaint description of the fiber fractured and detached while firing the laser was confirmed during evaluation of the returned complaint sample.The fracture location was determined to be just proximal from the marker band and where the physician was grasping the fiber.The fiber od was confirmed to meet product specification and there was no report that fiber was fractured when removed from the packaging.The likely root cause of this fiber fractured/detached event is handling damage to fiber during use, i.E.Fiber core fractured from physician's grip - resulting in laser energy escaping at the fractured site causing the spark, melted blue buffer layer and minor burn.Manufacturing personnel 100% visually inspect devices during the packaging process.This type of fiber kink/fracture damage would be noticed prior to shipment.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (14601411-01) which is supplied to the end user with the reported catalog number contains the following statement: warning contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your sales representative.Inspect prior to use to verify that no damage has occurred during shipping.Intended use the venacure evlt 400 m perforator and accessory vein ablation kit is intended for use in the treatment of superficial vein reflux of the greater saphenous vein associated with varicosities.The venacure evlt 400 m perforator and accessory vein ablation kit is indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for treatment of incompetent (i.E.Refluxing) perforator veins (ipvs).Equipment handling requirements venacure evlt 400 m perforator and accessory vein ablation kit: using sterile technique open the pack, place all contents into sterile field and inspect for damage.Do not use if any component is damaged.If damage is present, contact customer service or your local representative.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|
| |
|
Search Alerts/Recalls
|