• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BCI CAPNOCHECK II, HANDHELD CAPNOGRAPHY MONITOR; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BCI CAPNOCHECK II, HANDHELD CAPNOGRAPHY MONITOR; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 8400
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that "the unit was off per reading, re-calibrated".Patient involvement is unknown.
 
Additional Manufacturer Narrative
Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found no impact or corrosion damage was present and the tamper sticker was intact.Functional testing found the etco2 reading was approximately 45 mmhg @ 12 bpm.The specification at the current atmospheric pressure is 70 to 78 mmhg @ 12 bpm.Performed a hi/lo calibration and etco2 reading is still not within specification.Swapped out the co2 sensor and now getting a etco2 reading of approx.74 mmhg @ 12 bpm.The complaint was confirmed.Root cause was attributed to a faulty co2 sensor.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced faulty co2 sensor.Replaced naphion tubing and absorber as preventative maintenance.Device passed functional testing after the completed repairs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI CAPNOCHECK II, HANDHELD CAPNOGRAPHY MONITOR
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18352105
Report Number3012307300-2023-12086
Device Sequence Number8425341
Product Code CCK
UDI-Device Identifier10610586036736
UDI-Public(01)10610586036736(11)160126
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8400
Device Catalogue Number8400D
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/28/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2023
Supplement Date Received by Manufacturer01/15/2024
Initial Report FDA Received Date12/18/2023
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-