| Model Number |
3240121 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported event was unable to be confirmed due to limited information received from the customer.No device was returned to nuvasive qa for evaluation.Operative notes and/or medical records were not provided for review of usage/technique.A review of manufacturing records was unable to be performed as the lot information of the product involved in the event was not available.A definitive root cause was unable to be determined with the information provided.Labeling review: "pre-operative warnings: the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.Instruments should be protected during storage and from corrosive environments.Refer to cleaning and sterilization instructions below for all non-sterile parts.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the arm would not lock down, causing the retractor attachment to stay loose.There was no report of adverse impact to patient, user, or procedure.No additional information is available.
|
| |
|
Search Alerts/Recalls
|