| Catalog Number |
0279351400 |
| Medical Device Problem Code |
Electrical Shorting (2926)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: activation button stopped working, probable root cause: probe temperature extremely high.Probe short (possibly due to fluid ingress or other electrical failure).Use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event.
|
| |
|
Search Alerts/Recalls
|