| Model Number |
N/A |
| Medical Device Problem Codes |
Fracture (1260); Migration (4003)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Synovitis (2094); Subluxation (4525)
|
| Date of Event |
10/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: item number: unknown; lot #: unknown; item name: unknown oxford tibial component, item number: unknown; lot #: unknown; item name: unknown oxford femoral component.G2- switzerland.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient had an initial left medial unicondylar knee arthroplasty.Subsequently, the patient experienced a distortion trauma in which they sprained the left knee in a weight-bearing rotational movement.The patient underwent revision of the poly bearing due to unresolved pain in the knee.During the revision, the bearing was found luxated and fractured within the joint.Synovitis was also identified and removed.A new bearing was implanted without complications.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to correct the g3 in the previous medwatch report (3002806535-2023-00440).Correct g3: nov 25, 2023.
|
| |
|
Event or Problem Description
|
|
This report is being submitted to correct g3 in the previous medwatch report (3002806535-2023-00440).Correct g3: nov 25, 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.A review of the raw material certificate confirmed no abnormalities or deviations that could be related to the reported event.Medical records were provided and reviewed by a health care professional.The review identified a distortion trauma as contributing factor of the event: forced loaded flexion and rotation movement (weight-bearing maneuver up a staircase) - when turning, patient suffered a sprained trauma in his left knee, which resulted in a bang and he was no longer able to move his left knee properly.With the available information, a definitive root cause could not be determined.However, it is not excluded that the distortion trauma had an impact on the bearing fracture.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, d9, g3, g6, h3, h6, h11.Product has been returned and sent to our r&d department for analysis who reported as follows.The item is separated into (3) distinct pieces which were returned for examination.There is extensive (apparent) subsurface whitening on the lateral side and reaching to the posterior edge which may possibly indicate delamination.It is also reported that the posterior area of the bearing was "thinned out" upon explantation.Review of the complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and lot combination.Complaints are monitored per sop263 complaint trending process.With the available information, a definitive root cause could not be determined.However, the fracture of the bearing (which reportedly occurred with a "distortion trauma" of the knee when climbing stairs) was possibly due to overloading exacerbated by thinning and delamination of the bearing.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|