• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE 6.7/8.4 FR. X 430 MM, 5°, ANGLED OCULAR, 4.2 FR. CHANNEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE 6.7/8.4 FR. X 430 MM, 5°, ANGLED OCULAR, 4.2 FR. CHANNEL Back to Search Results
Model Number WA2UR11A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation was confirmed.There was an additional finding of scratch marks were visible inside the working channel near the distal end.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event or Problem Description
The customer reported to olympus that the ureteroscope 6.7/8.4 fr.X 430 mm, 5°, angled ocular, 4.2 fr.Channel tip looked displaced.The event occurred during a therapeutic lower ureteric stone procedure.The procedure was completed with a similar device.There was no patient harm associated with the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.Due to the light projectors emitting a large amount of energy, the connections of the endoscopic equipment and the distal end of the ureteroscope heat up.Due to the heat, the connection between the cover glass and the fiber tube has become loose.Cleaning agents and disinfectants may contain various aggressive components that can also damage the product.The following is included in the instructions for use: in order to remove residues, rinse the product thoroughly with deionized water just as described by the manufacturer of the cleaning agent.-do not use tap water for rinsing as it may be chlorinated.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE 6.7/8.4 FR. X 430 MM, 5°, ANGLED OCULAR, 4.2 FR. CHANNEL
Common Device Name
URETEROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18354279
Report Number9610773-2023-03684
Device Sequence Number12456271
Product Code FGB
UDI-Device Identifier04042761086387
UDI-Public04042761086387
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K200369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA2UR11A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer06/26/2024
07/08/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date06/27/2024
07/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-