| Catalog Number |
A897518 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that foley tray sticking together and sometimes leaked.
|
| |
|
Event or Problem Description
|
|
It was reported that foley tray sticking together and sometimes leaked.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be "compromised seal on package due to handling and/or processing issues".The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The product catalog number and the lot number for this device are unknown.Therefore, bd is unable to determine the associated labeling to review.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that foley tray sticking together and sometimes leaked.Per follow up via email on 29dec2023, it was reported that the item was #a897518.Defect was found prior to use and no patient was affected.The trays leaked lubricant sometimes.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be "compromised seal on package due to handling and/or processing issues".The dhr review could not be performed without a lot number.The product catalog number and the lot number for this device are unknown.Therefore, bd is unable to determine the associated labeling to review.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|