| Model Number |
453564423311 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The remote service engineer (rse) determined the cause of the reported problem was the transducer.The reported problem was confirmed.The rse determined that the transducer required replacement.The customer was provided a replacement transducer to resolve the issue.If additional information is received the complaint file will be reopened.
|
| |
|
Event or Problem Description
|
|
Philips received a complaint on the avcl-toco replacement xdr ism indicating that the cl transducer led turns cyan, with an intermittent audible alarm.It was reported that the avalon cl base did not detect this transducer, and there was no inop message on the monitor to state the battery on this transducer was empty.The device was in use at time of event, there was no adverse event reported.
|
| |
|