U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. THUNDERBEAT OPEN EXTENDED JAW; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. THUNDERBEAT OPEN EXTENDED JAW; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0920OE
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Pseudoaneurysm (2605); Hematemesis (4478); Melena (4480)
Date of Event 10/25/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the literature described that the device used was a "thunderbeat open jaw", olympus selected "tb-0920oe" as a representative product.The suspect device referenced in this report has not been returned to olympus.Additional information has been requested but no information could be provided.The literature is for additional information.This was translated to english electronically from its original japanese copy.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is received.
 
Event or Problem Description
Olympus medical systems corp.(omsc) received a literature titled "a case of ruptured gastroduodenal artery pseudoaneurysm 2 months after gastrectomy".Summary: the patient is an 83-year-old female.An open distal gastrectomy (d2 dissection, billroth i reconstruction) was performed for gastric cancer, and the patient was discharged from the hospital on the 15th postoperative day without any complications.The patient was diagnosed with pt2 (mp) n3bmo, pstage iiib, and was treated with postoperative adjuvant chemotherapy using ts-1 alone.On the 65th postoperative day, the patient suddenly developed hematemesis and melena, and contrast-enhanced ct revealed bleeding from a pseudoaneurysm in the gastroduodenal artery, so arterial embolization was performed.The patient was discharged from the hospital 30 days after embolization.Bleeding due to rupture of a pseudoaneurysm after gastrectomy is rare, and most cases occur secondary to suture failure or pancreatic fluid leakage within one month after surgery.On the other hand, there are also reports of arterial damage caused by the heat generated by devices such as ultrasonic coagulation and cutting devices, resulting in the formation of pseudoaneurysms.Perioperative management and appropriate device operation were considered important, keeping pseudoaneurysm rupture in mind as a complication of gastrectomy.There is no report of any olympus device malfunction in any procedure described in this study.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.This supplemental report includes additional information received from the author.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The author provided the following additional information: it is possible that the olympus device was related to the adverse event; however, it could not be determined.It cannot be verified, as the procedure was not reordered.From the nature of energy device, thermal effect and capitation can cause the adverse events described in the article.Thus, it cannot be ruled out that the thunderbeat was related.However, it was confirmed that an olympus device malfunction did not occur.
 
Additional Manufacturer Narrative
This report is being supplemented to provide the official english translation of the literature.Please see attached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT OPEN EXTENDED JAW
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18355868
Report Number9614641-2023-01952
Device Sequence Number12364105
Product Code GEI
UDI-Device Identifier04953170343315
UDI-Public04953170343315
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTB-0920OE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2023
Supplement Date Received by Manufacturer01/26/2024
01/30/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date01/26/2024
02/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age83 YR
Patient SexFemale
-
-