One opened suture, in a blister and in a pouch, in a bag was received for the report of shredded thread and white foreign material.The returned sample was visually inspected and found to be nonconforming, the suture was broken.No foreign material observed on the suture.A dimensional inspection for suture diameter and a functional test for suture tensile strength was performed and the sample was found to be conforming for both tests.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be non-conforming visually with a broken suture, however the sample was conforming for all functional and dimensional testing associated with the reported event, therefore shredded thread as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.The returned sample was found to visually conforming, therefore white foreign material as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the suture was manufactured to specifications.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|