| Catalog Number |
CS-17702-E |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: " the catheter was bent after the doctor inserted and pulled it out, cannot perform again.No patient harm.".
|
| |
|
Event or Problem Description
|
|
It was reported that: " the catheter was bent after the doctor inserted and pulled it out, cannot perform again.No patient harm.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Additional information states that the dilator was kinked/bent not the catheter as stated in the initial report.The report of a damaged dilator body was confirmed through complaint investigation.Visual analysis revealed that the distal portion of the dilator body contained a gradual and continuous bend.Despite the damage, the dilator met all relevant dimensional and functional requirements, and a device history record review did not reveal any relevant findings.Based on the customer report that the defect was observed "during use" and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: " the catheter was bent after the doctor inserted and pulled it out, cannot perform again.No patient harm.".
|
| |
|