Based on the information provided, it cannot be determined that the alleged bleeding event requiring cauterization is related to the activ.A.C.¿ ion progress¿ remote therapy monitoring system.The patient is on anticoagulation therapy, has a history of severe lower extremity arterial disease, and multiple other potentially contributing comorbidities.A device evaluation is pending return of the device.Device labeling, available in print and online, states: contraindications do not place foam dressings of the v.A.C.® therapy system directly in contact with exposed blood vessels, anastomotic sites, organs, or nerves.Warnings bleeding: with or without using v.A.C.® therapy, certain patients are at high risk of bleeding complications.The following types of patients are at increased risk of bleeding, which, if uncontrolled, could be potentially fatal: · patients who have weakened or friable blood vessels or organs in or around the wound as a result of, but not limited to: · suturing of the blood vessel (native anastomosis or grafts) / organ · infection · trauma · radiation · patients without adequate wound hemostasis · patients who have been administered anticoagulants or platelet aggregation inhibitors · patients who do not have adequate tissue coverage over vascular structures if v.A.C.® therapy is prescribed for patients who have an increased risk of bleeding complications, they should be treated and monitored in a care setting deemed appropriate by the treating physician.If active bleeding develops suddenly or in large amounts during v.A.C.® therapy, or if frank (bright red) blood is seen in the tubing or in the canister, immediately stop v.A.C.® therapy, leave dressing in place, take measures to stop the bleeding and seek immediate medical assistance.The v.A.C.® therapy units and dressings should not be used to prevent, minimize or stop vascular bleeding.Protect vessels and organs: all exposed or superficial vessels and organs in or around the wound must be completely covered and protected prior to the administration of v.A.C.® therapy.Always ensure that v.A.C.® foam dressings do not come in contact with vessels or organs.Use a thick layer of natural tissue should provide the most effective protection.If a thick layer of natural tissue is not available or is not surgically possible, multiple layers of non-adherent dressing material may be considered as an alternative, if deemed by the treating physician to provide a complete protective barrier.If using non-adherent materials, ensure they are secured in a manner that will maintain their protective position throughout therapy.Disclaimer: this information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a kinetic concepts, inc.Product malfunctioned, is defective or has caused serious injury.
|
|
On 24-nov-2023, the following information was provided to kci by the patient: on 20-nov-2023, the activ.A.C.¿ ion progress¿ remote therapy monitoring system was placed on hold allegedly due to bleeding from skin.On 01-dec-2023, the following information was provided to kci by the wound care center nurse: on (b)(6) 2023, the patient was seen at the wound care center and had active bleeding from the wound.Silver nitrate was used to cauterize the wound and v.A.C.® therapy was placed on hold.The cause of the bleeding was unknown, but patient is on anticoagulation medication.A device evaluation of the activ.A.C.¿ ion progress¿ remote therapy monitoring system is pending return of the device.
|
|
On 28-feb-2024, a device evaluation was completed by kci quality engineering.On (b)(6) 2023, the device was tested per quality control procedure by a kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2023, the device was placed with the patient.On (b)(6) 2024, the device was tested per quality control procedure by a kci service center and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit.
|