| Catalog Number |
151640706 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tab on the poly broke while inserting and it would no longer lock.Surgical delay was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : it was reported that the tab on the poly broke while inserting and it would no longer lock.Surgical delay was unknown.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection of the returned sample revealed that the tab of the attune ps fb insrt sz 7 6mm was found fractured and with marks along one of the posterior grooves that is designed to slide into the locking feature of the mating tibial tray.The observed damage suggest unsuccessful insertion attempts during the surgical process, however a definitive root cause is not possible to be determined.The mating tibial tray component was not returned, therefore, a functional test was not able to be performed.However, due to the visual damage observed.The difficulty/inability to assemble mating devices can be confirmed.The attune knee system surgical technique (dsus/jrc/0316/1437 rev.K) advises on pages 79-80, ¿for fixed bearing tibial components, angle the tibial insert posteriorly and slide the posterior tabs into the posterior undercuts of the tibial base.The fixed bearing tibial insert is impacted into place on the tibial base, using the fixed bearing insert impactor.Position an impactor at approximately 60 degrees on the insert so that the notch rests on the anterior edge of the center of the insert.Use a mallet to strike the fixed bearing insert impactor.¿ following these steps will successfully locking the insert into the base as intended.The observed damaged condition suggests the posterior grooves on the underside of the tibial insert were not properly aligned with the undercut locking features of the tibial base.This would contribute to the difficulty/inability to mate the insert with the tibial component following multiple insertion attempts.The overall complaint was confirmed as the observed condition of the attune ps fb insrt sz 7 6mm would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|