• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FB INSRT SZ 7 6MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FB INSRT SZ 7 6MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151640706
Medical Device Problem Codes Fracture (1260); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the tab on the poly broke while inserting and it would no longer lock.Surgical delay was unknown.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : it was reported that the tab on the poly broke while inserting and it would no longer lock.Surgical delay was unknown.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection of the returned sample revealed that the tab of the attune ps fb insrt sz 7 6mm was found fractured and with marks along one of the posterior grooves that is designed to slide into the locking feature of the mating tibial tray.The observed damage suggest unsuccessful insertion attempts during the surgical process, however a definitive root cause is not possible to be determined.The mating tibial tray component was not returned, therefore, a functional test was not able to be performed.However, due to the visual damage observed.The difficulty/inability to assemble mating devices can be confirmed.The attune knee system surgical technique (dsus/jrc/0316/1437 rev.K) advises on pages 79-80, ¿for fixed bearing tibial components, angle the tibial insert posteriorly and slide the posterior tabs into the posterior undercuts of the tibial base.The fixed bearing tibial insert is impacted into place on the tibial base, using the fixed bearing insert impactor.Position an impactor at approximately 60 degrees on the insert so that the notch rests on the anterior edge of the center of the insert.Use a mallet to strike the fixed bearing insert impactor.¿ following these steps will successfully locking the insert into the base as intended.The observed damaged condition suggests the posterior grooves on the underside of the tibial insert were not properly aligned with the undercut locking features of the tibial base.This would contribute to the difficulty/inability to mate the insert with the tibial component following multiple insertion attempts.The overall complaint was confirmed as the observed condition of the attune ps fb insrt sz 7 6mm would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 7 6MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18356373
Report Number1818910-2023-25572
Device Sequence Number15177378
Product Code JWH
UDI-Device Identifier10603295050216
UDI-Public10603295050216
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number151640706
Device Lot NumberD23060666
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/12/2023
Supplement Date Received by Manufacturer01/18/2024
02/09/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date01/18/2024
02/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-