| Catalog Number |
07154003001 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation of an issue with cobas infinity core software version 3.03.17.The error code "3428" sent by the instrument was captured as a patient result for the ggt2 test.This "result" was manually validated by the customer and released.
|
| |
|
Additional Manufacturer Narrative
|
|
The instrument connectivity agent (ica) is the module that communicates with instruments connected to cobas infinity core software.Its function is to exchange information with any instrument connected to the laboratory information system (lis).To do so, it is necessary to configure the infinity software to the corresponding driver to connect with the instrument.In this case, the customer is using the alinity ci driver version 3.0.0.29.The investigation determined the driver is not correctly identifying "code error" messages and treats them as regular patient results.This causes the error code number sent by the instrument to be saved as a patient test result in the cobas infinity software.It was found that the message included the "x" flag indicating the test result could not be obtained; this flag was correctly added to the result in the cobas infinity software.The investigation confirmed this as a software issue.A workaround has been provided.A new driver version (alinity ci driver version 3.0.1.33) has been released to resolve the issue.
|
| |
|
Search Alerts/Recalls
|