| Catalog Number |
150401106 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526); Unspecified Tissue Injury (4559)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient had a complete knee replacement including a 38mm patella attune knee in (b)(6) 2019.It had to be revised in 2021 and again in 2023.On (b)(6) 2021, the patient had a left knee revision to address pain, instability and ¿general loosening prosthesis¿.Pathology reports from the surgery noted the patient had tissue reaction from polyethylene debris.The tibial tray, insert, and femoral component were removed and competitor components were implanted during this procedure.The patella was left in-situ.(b)(4).In 2023, the patient was revised to address pain, giant cell reaction, tissue reaction due to polyethylene debris.Only the depuy patella was revised.Competitor patella was implanted along with competitor cement during this procedure.(b)(4).Doi: (b)(6) 2019 dor: 2023 left patella only.Doi: (b)(4).2019.Dor: (b)(4) 2021.Affected side: left.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).E3: the initial reporter has been removed for confidentiality/privacy.The initial reporter is the patient.H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|