| Model Number |
0600-0767 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer declined the option to have the reported glidescope core quickconnect cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Review of complaint history found that this customer reported one similar incident using a different glidescope core quickconnect cable.Follow-up information received from the customer's verathon territory manager reported that this most recent incident ocurred after replacing their initial glidescope core quickconnect cable and using a different generation of bflex single-use bronchoscope with a loaner glidescope core 15-inch monitor.The customer was unable to isolate the issue to a particular system component.The customer's verathon territory manager reported that the customer will be replacing their glidescope core system following the reported event.Trending analysis for the glidescope core quickconnect cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized by verathon.Corrective action is currently not required at this time.Verathon will continue to monitor for any ongoing trends.
|
| |
|
Event or Problem Description
|
|
A customer reported that during a patient procedure, using a glidescope core quickconnect cable, black and grey scrambled lines appeared on the connected glidescope core 15-inch monitor with a bflex 2 large 5.8 single-use bronchoscope.The procedure was completed using a backup glidescope system but the same bflex 2 large 5.8 single-use bronchoscope which was made available in an unspecified amount of time.No delay in the procedure or harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only**.This is a supplemental report correcting the unique device identifier (udi) information in section d4.In the initial report submission, the udi information field was left blank due to the subject device lot number and manufacture date being unknown.Despite verathon being unable to obtain the lot information from the customer, this supplemental report provides the udi-di for the subject device model.
|
| |
|