| Model Number |
0998-UC-3023-65 |
| Medical Device Problem Codes |
No Display/Image (1183); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during daily inspection, the cs300 intra-aortic balloon pump (iabp) unit screen display failed.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The getinge field service engineer (fse) replaced the video receiver board and display cable.Operational inspection was performed based on inspection records.The unit is in good condition.
|
| |
|
Event or Problem Description
|
|
It was reported that during daily inspection performed by getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) unit screen display failed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to character restrictions in block e1 event site name : (b)(6).Due to character restrictions in block e1 city : (b)(6).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The following investigation was performed by technician of the maquet failure analysis and testing dept.(fat) wayne, nj.The failure analysis and testing dept.Received the following parts associated with this complaint: pcb, color video receiver serial number (b)(6).Cable, video receiver to lcd 08 30_meritec.These parts were received with a reported unit failure of screen display failure.The failure analysis and testing dept.Installed the parts into the cs300 test fixture and tested parts to factory specifications per the cs300 service manual.After an extended run-time the fat could not duplicate the reported complaint by the customer.Pcb, color video receiver serial number (b)(6) and cable, video receiver to lcd 08 30_meritec passed testing.The non-conformances with the returned components were not identified.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** providing updated device identification information in alignment with gudid (d1, d2, d3, d4, g4, h4, h5).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|