According to the reporter, during a laparoscopic cholecystectomy, during insertion of the trocar in the patient, the suture was in place, the white part was squeezed to put the cannula and it broke.It was noted that the white wings where the suture was tied was broken.A new trocar was used to resolve the issue.There was no patient injury.
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.The visual inspection of the device noted that the spring grip for the threaded anchor was broken.The obturator was received inserted into the cannula, prolonged insertion can cause the seals to become stretched.The obturator was removed and the duckbill seal was stretched out.The cannula, obturator and stopcock appeared intact.Functional testing noted that the device failed an air leak test.The product was shipped back with the obturator inserted in the cannula and therefore the length of time it was inserted during shipment may have also resulted in the stretched seals.The spring grip was evaluated, a proper weld was noted.It was reported that the spring grip and the anchor tabs were broken.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.These issues can occur if excessive force was applied during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use special care when introducing or removing sharp-edged or sharp-angled endoscopic instruments to minimize the potential of inadve rtent damage to the seal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|