MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION GLIDESHEATH SLENDER; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION GLIDESHEATH SLENDER; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that the glidesheath slender sheath was leaking from the hub.The event occurred intra-operative.There is not a direct allegation that the reported device caused or contributed to patient injury and/or need for medical intervention.
 
Additional Manufacturer Narrative
A1: patient identifier: requested, not provided due to hospital policy.A2: age & date of birth: requested, not provided due to hospital policy.A3: patient sex: reqeusted, not provided due to hospital policy.A4: weight: requested, not provided due to hospital policy.A5: ethnicity: requested, not provided due to hospital policy.A6: race: requested, not provided due to hospital policy.D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.E3: occupation: supervisor, ccl.The actual device was not available; therefore, the actual device will not be returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide additional information to sections b5, h6: clinical code, and h6: impact code.
 
Event or Problem Description
Additional information was received on 14dec2023: the procedure performed was a cardiac catheterization.The leakage was unresolved.The leakage of blood was less than 250cc's.The patient was in stable condition.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for valve leakage because the complaint sample was not returned for assessment.The exact root cause could not be determined.Historically, valve leakages have been caused by off-center insertion of the dilator.Review of the device history record (dhr) showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the design failure mode and effects analysis (dfmea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESHEATH SLENDER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
gina digioia
950 elkton blvd
elkton, MD 21921
4103927218
MDR Report Key18357672
Report Number1118880-2023-00455
Device Sequence Number6407797
Product Code DYB
UDI-Device Identifier00389701011950
UDI-Public00389701011950
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K173831
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2026
Device Model NumberN/A
Device Catalogue Number60-1050
Device Lot Number0000402602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2023
Supplement Date Received by Manufacturer12/14/2023
02/09/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date01/10/2024
03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1