| Model Number |
100014514 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One ensite velocity¿ amplifier was received for evaluation.Based on the information provided to abbott and the investigation performed, the reported event was able to be confirmed, and the root cause was attributed to a voltage out of range condition during post for catheter amplifier board in slot 9.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|
Event or Problem Description
|
|
During the atrial fibrillation procedure, self-test failure issues with the amplifier required multiple troubleshooting measures which did not resolve the issue and the procedure was cancelled.It was noted that the amplifier booted up and connected to the dws without issue, however, after a few moments, the amplifier status light went to solid amber.The amplifier was rebooted several times, all connections were unplugged, the power was unplugged and let sit for a minute all without resolution.Rebooted again to solid amber and it remained solid amber.The case was cancelled with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d1, d4, g3, h2, h11.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.Additional information: d1, d4, g3, h2, h11.
|
| |
|