Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during set up the truepass suture passer broke.There was a back up device available and no surgical delay was reported.No patient involvement was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.A visual inspection revealed it is not in its original packaging.The device shows wear from use with the device identification markings degraded but legible.The upper jaw is fractured and the torsion spring and suture capture door on the distal end of the shaft are fractured away.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include an application of unintended inappropriate or excessive force to the device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|