As reported by the equinoxe shoulder study, the patient, (b)(6), had an initial left tsa on (b)(6) 2017.Patient (b)(6) presented on (b)(6) 2017 with instability described as having subluxation that are self-reduced.The case report form indicates that this event is possibly related to device and definitely related to procedure.Outcome is continuing; patient advised to schedule revision (poly exchange).Due to clinical study, no devices will be returned for evaluation.
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: a, b, c, d, e, f, g.No explant date.The cause of the patient¿s shoulder dislocation and subsequent revision reported cannot be conclusively determined; however, it may be due to soft tissue imbalance, patient anatomy, implant positioning, and/or implant selection.Dislocation is a known risk, as outlined in the ifu.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|