• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGIPRO II; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGIPRO II; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLEN Back to Search Results
Model Number VP556X-2
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: vp706x-2, srgproii 6-0 30 bl cv-11 da vp706x-2 (lot#d1c0719y) vp706x-2, srgproii 6-0 30 bl cv-11 da vp706x-2 (lot#d2l2537y); medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, both sutures were accidentally cut during the surgeon's attempt to repair the arteries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIPRO II
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLEN
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18358704
Report Number9612501-2023-02421
Device Sequence Number12766174
Product Code GAW
UDI-Device Identifier10884521758582
UDI-Public10884521758582
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K010909
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberVP556X-2
Device Catalogue NumberVP556X-2
Device Lot NumberD0K1796Y
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2023
Initial Report FDA Received Date12/19/2023
Date Device Manufactured10/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient SexUnknown
-
-