| Model Number |
S602 |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker did not pace when required, ventricular pacing was delayed.Technical services (ts) was consulted and reviewed possible reasons for the exhibited behavior.This pacemaker remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker did not pace when required, ventricular pacing was delayed.Technical services (ts) was consulted and reviewed possible reasons for the exhibited behavior.This pacemaker was explanted due to normal battery depletion after approximately ten years implanted.No adverse patient effects were reported.
|
| |
|