Model Number KD-VC611Q-07201S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has been returned to olympus for evaluation and the investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported that the sphincterotome cutting wire broke at the very end of an unspecified therapeutic procedure which was completed with the same device.There was no patient or user impact or additional operating time due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.The cutting wire was broken.The broken portion was scorched and melted.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a likely mechanism causing the broken cutting wire might be the following: 1.The device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.2.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were being close to each other.3.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.4.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.The event can be detected/prevented by following the instructions for use which state: - "since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the wire is very short, the output is too high or activated while the wire touches metal parts of the endoscope, or the wire is tightened too strong.When the wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the wire will be pushed out toward the papilla or move sideways.If the wire breaks off, stop the output immediately and pull the slider completely to retract the broken wire into the tube.Then withdraw the instrument from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct and/or damage of the endoscope could result.- be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.- do not activate output while the cutting wire touches the metal parts of the endoscope, or they are being close together.This could burn the tissue and/or damage the endoscope or the instrument." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED)
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18359779
Report Number9614641-2023-01954
Device Sequence Number15431868
Product Code KNS
UDI-Device Identifier04953170399381
UDI-Public04953170399381
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141991
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-VC611Q-07201S
Device Lot Number31V 06
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer03/05/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-