• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS INTEGRA GLUCOSE HK GEN. 3; HEXOKINASE, GLUCOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS INTEGRA GLUCOSE HK GEN. 3; HEXOKINASE, GLUCOSE Back to Search Results
Catalog Number 04404483190
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas 6000 c 501 serial number was (b)(6).The investigation is ongoing.
 
Event or Problem Description
There was an allegation of a questionable glucose (gluc hk gen.3) result from the cobas 6000 c 501.The initial result was 9 mg/dl with a data flag and the repeat result was 72 mg/dl.The repeat result from another cobas c501 was 74 mg/dl.The questionable result was not reported outside of the laboratory.The repeat result from the other analyzer was believed correct.
 
Additional Manufacturer Narrative
The field service engineer checked the analyzer and did not find any issues.He ran an assay performance check and the customer performed calibration and qc.The customer noted that is issue had occurred with a very short sample.Therefore, the issue was consistent with a preanalytical handling issue by the customer.The investigation did not identify a product problem.The specific cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INTEGRA GLUCOSE HK GEN. 3
Common Device Name
HEXOKINASE, GLUCOSE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18359804
Report Number1823260-2023-04115
Device Sequence Number20811505
Product Code CFR
UDI-Device Identifier04015630920297
UDI-Public04015630920297
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K061048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04404483190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2023
Supplement Date Received by Manufacturer01/04/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date01/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
Patient SexFemale
-
-