• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ENGAGE POROUS TIBIAL TRAY SZ 4-RT MED; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ENGAGE POROUS TIBIAL TRAY SZ 4-RT MED; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 1-10012-450
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 11/29/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference number: (b)(4).H10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Event or Problem Description
It was that, after a uka had been performed on (b)(4) 2023, the patient underwent a revision surgery to a tka on (b)(4) 2023, due to the engage porous tibial tray sz 4-rt med not being ingrown and had startup pain.In addition, one (1) engage tibial insert sz 4-rt med 9mm, one (1) engage porous femoral sz 5-rt med and one (1) engage porous femoral sz 5-rt med were removed.
 
Additional Manufacturer Narrative
The reported event was analyzed.Although the involvement of all the devices was reported, the relationship with the investigated failure was directly associated to the engage porous tibial tray.Therefore, no investigation is deemed for the other devices.The device was not returned for evaluation.However, the photographs were reviewed, and revealed tissue residue and blood, however, there are no visible defects that can be associated with the reported adverse event.The clinical/medical investigation concluded that, based on the limited information provided, the clinical root cause of the reported pain is likely due to inadequate osseointegration of the tibia component; however, the root cause of the failed tibia component osseointegration cannot be definitively concluded.The patient impact is the reported pain, and the revision of the unicompartmental knee arthroplasty to total knee arthroplasty.The patient¿s current health status is unknown.A review of the production record did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for engage partial knee system revealed in adverse effects and complications that pain can result from improper positioning or loosening of the components.There is not enough data available to conclude that the overall clinical benefit outweighs the potential risk profile when compared to the state of the art.The existing data identifies a potential signal that the performance is an outlier vs the state of the art with respect to the risk for revision.In addition, a historical review concluded that no previous escalated actions for this type of issue were identified.However, as a correction action a voluntary market removal will be performed for the engage cementless partial knee system due recent complaint data that indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include abnormal motion over time, bone degeneration, lack of ingrowth, osteolysis and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENGAGE POROUS TIBIAL TRAY SZ 4-RT MED
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18359968
Report Number1020279-2023-02528
Device Sequence Number7771771
Product Code HSX
UDI-Device Identifier00885556873182
UDI-Public885556873182
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K190439
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/13/2026
Device Catalogue Number1-10012-450
Device Lot Number100267
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer01/17/2024
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date01/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1-10003-500/ENGAGE POROUS FEMORAL SZ 5-RT MED.; 1-10011-200/ENGAGE TIBIAL ANCHOR STEM SZ 3-4.; 1-10013-459/ENGAGE TIBIAL INSERT SZ 4-RT MED 9MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexUnknown
Patient Weight113 KG
-
-