• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 21AGFN-756
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Codes Aortic Valve Stenosis (1717); Respiratory Insufficiency (4462)
Date of Event 11/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2023, a 21mm sjm regent heart valve with flex cuff was selected for an implant procedure in a patient with prior aortic stenosis and respiratory distress.It was noted during procedure that the valve had been oversized and would not enter the annulus fully, but the valve had already been sutured.It was noted that the event resulted in exacerbation of the patient's aortic stenosis and respiratory distress.The decision was made to remove and replace the 21mm sjm regent heart valve with flex cuff, with a 21mm epic supra valve to complete the procedure.There was no other treatment performed due to the patient's exacerbated aortic stenosis and respiratory distress.The patient was not taken off bypass due to this event.The patient was reported as stable at the time of report.
 
Additional Manufacturer Narrative
The reported event was due to exacerbation of the patient's aortic stenosis and respiratory distress.It was also reported that the valve was oversized and would not enter the annulus fully, but the valve had already been sutured.The valve was explanted and no other treatment was performed.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18360196
Report Number2135147-2023-05592
Device Sequence Number18085420
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number21AGFN-756
Device Lot NumberC00001028
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2023
Supplement Date Received by Manufacturer12/28/2023
Initial Report FDA Received Date12/19/2023
Supplement Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-