| Model Number |
A2025-200 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation, and before use in the patient, when negative was pulled on the armada 14 balloon, there was a hole [tear] in the balloon and air bubbles continued to fill.A new armada was used to continue the procedure without further issue.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 correction: model number entered.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: updated the device identification fields to align with the information in the corresponding fields in gudid.
|
| |
|
Additional Manufacturer Narrative
|
|
D1 correction: updated device brand name to align with the information in the corresponding fields in gudid.
|
| |
|