| Model Number |
HUMIDIFIER EMPTY 6 PSI BLUE CAP 50/CS |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
09/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to vyaire medical that after attempting to place a patient on an intermediate flow nasal cannula (12 lpm).The patient desaturated to 77% spo2.It was discovered that the 002006 humidifier empty 6 psi blue cap 50/cs had no hole present on the 02 flow outlet.The respiratory therapist then placed the patient on hfnc (high flow nasal cannula) at 50% 02.At this time, the customer has not provided any information regarding patient harm or injury associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device is not available for evaluation, however, we request a device history record review.The root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|