| Model Number |
232-10-85-250 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The top portion of the graft bolt has not returned for evaluation.The identifying lot number was not provided.The root cause could not be determined.If additional information is provided, a supplemental report will be submitted.Labeling review: "warnings/cautions/precautions: care should be taken in performing screw hole preparation to facilitate a proper graft bolt insertion trajectory and implantation.Confirm under fluoroscopy that the graft bolt insertion angle is as close as possible to a 40° trajectory.A shallow or incorrect graft bolt trajectory may result in encroachment of the spacer and lead to graft bolt breakage.".
|
| |
|
Event or Problem Description
|
|
During an anterior lumbar interbody fusion procedure, the graft bolt broke in two when the surgeon was tightening.The head of the screw was recovered, but the shank remains in situ.There are no plans for revision surgery.Additional information was requested and is not available.
|
| |
|