| Model Number |
HUMIDIFIER KIT 500ML 12/CS |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
10/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that 2620 humidifier kit 500ml 12/cs is producing 0 air sound from the cannula even when it is on 4 liters during patient use.The patients saturation dipped below 60.The nurse disconnected the oxygen tubing from the humidifier and replaced the humidifier with a new one and the patients saturation increased to 78.They were able to wean 02 back down to 2l.They observed a large crack in the plastic where oxygen must have been leaking out.Furthermore, there was no harm on the patient.
|
| |
|