• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INC AIRLIFE¿ STERILE WATER FOR INHALATION, USP, 500 ML; HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INC AIRLIFE¿ STERILE WATER FOR INHALATION, USP, 500 ML; HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) Back to Search Results
Model Number HUMIDIFIER KIT 500ML 12/CS
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Date of Event 10/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
It was reported to vyaire medical that 2620 humidifier kit 500ml 12/cs is producing 0 air sound from the cannula even when it is on 4 liters during patient use.The patients saturation dipped below 60.The nurse disconnected the oxygen tubing from the humidifier and replaced the humidifier with a new one and the patients saturation increased to 78.They were able to wean 02 back down to 2l.They observed a large crack in the plastic where oxygen must have been leaking out.Furthermore, there was no harm on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRLIFE¿ STERILE WATER FOR INHALATION, USP, 500 ML
Common Device Name
HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)
Manufacturer (Section D)
VYAIRE MEDICAL INC
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
2402760001
MDR Report Key18360734
Report Number3013421741-2023-00051
Device Sequence Number12766053
Product Code BTT
UDI-Device Identifier10190752141643
UDI-Public(01)10190752141643(10)TL2301009
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K853146
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHUMIDIFIER KIT 500ML 12/CS
Device Catalogue Number2620
Device Lot NumberTL2301009
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Initial Report FDA Received Date12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-