Model Number 26924
Device Problem Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2023
Event Type  malfunction  
Event Description
It was reported that stent dislodgement occurred.The target lesion was located in the common iliac artery.An 8.0x40x75cm express ld stent was advanced for treatment.During the procedure, the stent came off from the balloon prior to deployment.An 8 x 60 x 130 innova vascular stent was used to trap the express ld stent up against the wall vessel.However, the innova stent did not reach the desired target area for placement.The physician plans to send back the patient to stent the intended area.No complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18360742
MDR Text Key331207299
Report Number2124215-2023-71829
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08714729874188
UDI-Public08714729874188
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26924
Device Catalogue Number26924
Device Lot Number0031095145
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-