|
The device was returned to olympus for evaluation, and the evaluation findings were as follows: the bending section cover had a pinhole, the hole was taped, and was leak tested again; with no other leaks found.As per the advanced diagnostics (ad) process, the bending section cover, hole/cut was repaired.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation the root cause could not be identified; however, it is likely that an abnormality of electrical parts in the subject device occurred due to water invasion, leading to the malfunction.The event can be prevented by following the instructions for use which state: (same as bf-mp160f) important information ¿ please read before use precaution for disappeared or frozen endoscopic image; warning ·follow the precautions given below when handling the instrument.Otherwise, the endoscopic image may disappear unexpectedly, or the frozen image may not be restored during the examination.Precaution for disappeared or frozen endoscopic image; caution ¿ turn the video system center off before connecting or disconnecting the videoscope cable from the electrical connector on the endoscope.Turn the switch on or off only when the videoscope cable is connected to both the video system center and electrical connector on the endoscope.Failure to do so can result in equipment damage, including destruction of the ccd.¿ do not touch the electrical contacts inside the electrical connector.Ccd damage may result.3.6 inspection of the endoscopic system inspection of the endoscopic image 4.While observing the palm of your hand, confirm that the wli and nbi endoscopic image is free from noise, blur, fog, or other irregularities.5.Angulate the bending section and confirm that the endoscopic image is free from momentary disappearing or other irregularities.Chapter 5 troubleshooting if the endoscope is visibly damaged, does not function as expected or is found to have irregularities during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope.Contact olympus.Some problems that appear to be malfunctions may be correctable by referring to section 5.1, ¿troubleshooting guide¿.If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to olympus for repair.Olympus does not repair accessory parts.If an accessory part becomes damaged, contact olympus to purchase a replacement.Olympus will continue to monitor field performance for this device.
|