• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVENU MEDICAL ELLIPSYS CATHETER; PERCUTANEOUS CATHETER FOR CREATION OF AN ARTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVENU MEDICAL ELLIPSYS CATHETER; PERCUTANEOUS CATHETER FOR CREATION OF AN ARTER Back to Search Results
Catalog Number AMI6005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 07/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During index procedure a ellipsys catheter was used to treat a lesion located in the right arm.Approximately 1 week post procedure the patient underwent a secondary procedure during which a medtronic balloon was used to treat the anastomosis.12 days post index procedure patient suffered right arterial anastomosis stenosis.Patient presents for follow up with no thrill or bruit of right proximal forearm av fistula.Fistulagram completed on same day reveals 80% stenosis of the right proximal radial artery stenosis of the area between the bifurcation and and the anastomosis.Patient was treated using a medtronic balloon with angioplasty of the arterial anastomosis/perforator vein and of the proximal part of the radial artery, resulting in improved flow and low resistance waveform.Week 4 follow up dus showed patent anastomosis with a flow of 0.64 cm/s and target vessel flow of 649.9 ml/min, indicating resolution of right arterial anastomosis stenosis. patient recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSYS CATHETER
Common Device Name
PERCUTANEOUS CATHETER FOR CREATION OF AN ARTER
Manufacturer (Section D)
AVENU MEDICAL
27123 calle arroyo
#2101
san juan capistrano CA 92675
Manufacturer (Section G)
AVENU MEDICAL
27123 calle arroyo
#2101
san juan capistrano CA 92675
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18361716
Report Number3014251914-2023-00024
Device Sequence Number20776175
Product Code PQK
UDI-Device Identifier00763000496975
UDI-Public00763000496975
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K191114
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/20/2024
Device Catalogue NumberAMI6005
Device Lot Number001007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Initial Report FDA Received Date12/20/2023
Date Device Manufactured04/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight64 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-