• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MILAGRO ADVANCE SCREW 8X30MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MILAGRO ADVANCE SCREW 8X30MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 231821
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.H4: the device manufacture date was unknown.D10: concomitant devices: 2x truespan 24 degree peek, therapy date: 12/12/2023.Udi: (b)(4).
 
Event or Problem Description
It was reported by the sales rep that during a knee surgery on 12/12/2023 the truespan device x2 could not fire.Another like device was used to complete the procedure.The milagro advance screw 8x30mm device anchor was broken off when the anchor was inserted, removed all the broken parts.Another like device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.This is report 2 of 3 of the same event.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the investigation has been updated to reflect the correct information: investigation summary the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number (8l58119), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Correction h4: the device manufacture date was reported as unknown on the initial report.Please note that the date of manufacture has been updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MILAGRO ADVANCE SCREW 8X30MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INT MITEK
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 
3035526892
MDR Report Key18362439
Report Number1221934-2023-04675
Device Sequence Number8014803
Product Code HWC
UDI-Device Identifier10886705022267
UDI-Public10886705022267
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K123362
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Catalogue Number231821
Device Lot Number8L58119
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2023
Supplement Date Received by Manufacturer12/20/2023
Initial Report FDA Received Date12/20/2023
Supplement Report FDA Received Date01/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-