|
SYNTHES TRAUMA : CROSSROADS EXTREMITY SYSTEMS, LLC MINIBUNION DRILL KIT - STERILE; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Catalog Number |
3100-0001 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure on (b)(6) 2023, the holes in the capital fragment control guide were found to be too small for the pins provided.This same issue was found on both guide kits.The procedure was completed successfully by using a smaller wire to pass through the capital fragment guide.This report is for a minibunion drill kit ¿ sterile.This is report 3 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3,h4,h6 the product was not returned to depuy synthes, however photos were provided for review.The photo investigation cannot confirm the device issue reported in minibunion drill kit, sterile per it is only visible a k-wire instrument.Photos don't provide enough evidence to confirm the alleged reported condition.Functionality of the device cannot be assessed through photo evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the minibunion drill kit, sterile would not contribute to the complained device issue.Based on the investigation findings it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H4,h6 part: 3100-0001, synthes lot: 502410, supplier lot # n/a, release to warehouse date: june 22, 2023, manufactured by: crossroads extremity systems, llc, no non conformance reports were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Photo investigation: the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed that it is visible the k-wire does not fit through the device guide holes.This is due to the condition of the guide, therefore, the condition can be confirmed root cause was determined to be during manufacturing process, where the wrong cycle program for the drying oven was selected and warping occurred.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the minibunion drill kit, sterile would contribute to the complained device issue.Based on the information currently available, a product issue was identified during the investigation.This product issue will be addressed through j&j medtech orthopaedics quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr); part: 3100-0001; synthes lot: 502410; release to warehouse date: june 22, 2023.Manufactured by: crossroads extremity systems, llc.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|