• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL CZECH REPUBLIC A. S BLU SELECT SUCTIONAID CUFFED TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL CZECH REPUBLIC A. S BLU SELECT SUCTIONAID CUFFED TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/870/090
Medical Device Problem Codes Defective Component (2292); Gas/Air Leak (2946)
Health Effect - Clinical Codes Aspiration Pneumonitis (4455); Insufficient Information (4580)
Date of Event 11/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3: month and year of event have been provided, day is unknown.E2: initial reporter - head of speech therapy.H3 - other: device is not available to be returned for investigation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the device was malfunctioning during use.The issues include the malfunctioning of the subglottic suction valve on the newly inserted tracheostomy tube.Additionally, the cuff of the tracheostomy tube exhibited immediate leakage after insertion.This prompted the need for multiple tube changes.This situation is of particular concern as the patient undergoing these changes is at high risk of aspiration.It was reported that the following measures were taken in order to "solve the problem: tk change, survey of laboratory parameters and antibiosis." there was patient involvement.The samples are not available for evaluation.The products were found to be defective between 1-nov-2023 and 14-nov-2023.Only one patient was involved.(patient identifier (b)(6).
 
Additional Manufacturer Narrative
Device evaluation: a review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No trend of confirmed complaints in relation with this issue was identified.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLU SELECT SUCTIONAID CUFFED TRACHEOSTOMY TUBE
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL CZECH REPUBLIC A. S
olomoucka 306
hranice 753 0 1
EZ  753 01
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18365599
Report Number3011237704-2023-00005
Device Sequence Number7806364
Product Code BTO
UDI-Device Identifier15019517076240
UDI-Public(01)15019517076240(17)271026(10)4303381
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101/870/090
Device Catalogue Number101/870/090CZ
Device Lot Number4303381
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2023
Supplement Date Received by Manufacturer01/17/2024
Initial Report FDA Received Date12/20/2023
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-