| Model Number |
24657 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Premature Activation (1484); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the stent partially deployed.This 7x40, 130 cm eluvia drug-eluting vascular stent system was selected for use in a superficial femoral artery angiogram procedure.Difficulty was experienced during insertion of the stent system at the hub of the 6fr sheath.Then it was noticed that the stent had partially deployed, preventing the system from being introduced into the sheath.The device was removed, and a second stent of the same size was successfully deployed.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to b5: describe event or problem.Correction to h6: device codes.
|
| |
|
Event or Problem Description
|
|
It was reported that the stent inadvertently deployed.This 7x40, 130 cm eluvia drug-eluting vascular stent system was selected for use in a superficial femoral artery angiogram procedure.Difficulty was experienced during insertion of the stent system at the hub of the 6fr sheath.Then it was noticed that the stent had inadvertently deployed, preventing the system from being introduced into the sheath.The device was removed, and a second stent of the same size was successfully deployed.There were no patient complications.
|
| |
|