| Model Number |
09007733190 |
| Medical Device Problem Codes |
High Test Results (2457); Chemical Problem (2893)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The qc recovery data provided was acceptable.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys t3, elecsys t4 assay, and elecsys tsh assay results for 6 patient samples on a cobas e 801 analytical unit compared to tree competitor analyzers.This medwatch will cover t3.Refer to medwatch with a1 patient identifier (b)(6) for information on the t4 results and medwatch with a1 patient identifier (b)(6) for information on the tsh results.Refer to the attachment to the medwatch for all patient data.The units of measurement for the tsh results and for the competitor analyzer results were not provided.The reference ranges for all assays were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d4 reagent lot number has been updated.The customer's t3 reference range is 80-200 ng/dl.The customer stated that their qc recovery data was acceptable.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event was found to be consistent with pre-analytical sample handling issues.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** d3 is the initial distributor in the us.
|
| |
|