• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 04837738001
Medical Device Problem Codes Image Display Error/Artifact (1304); Misassembly by Users (3133)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
We received an allegation about a faded display when a patient tested with a coaguchek xs meter.The reporter alleged that since the meter was flickering, he replaced the batteries.He also alleged that the meter did not power on after that.The reporter was advised to turn off the meter and press the m button to access the meter memory.The meter was powered on with an error message.The reporter alleged that he could not determine if the error message was an error 4 or an error 9 as the numbers were faded.
 
Additional Manufacturer Narrative
The meter was requested for investigation.The reporter stated that the original batteries were not leaking.The investigation is ongoing.
 
Additional Manufacturer Narrative
Sections d9 and h3 were updated.The meter was provided for investigation.The error log could not be read out as the device did not turn on during the investigation.The meter was disassembled for further investigation.The optical inspection of the printed circuit board revealed contamination due to residues of fluid ingress (leaked battery acid, cleaning solution).The residues are visible in the area on the printed circuit where communication with the display takes place.The contamination on the battery contacts led to the mentioned problem with the faded segments and was the cause of the power interruption.The investigation determined there was a user error.The root cause was due to contamination of the contacts due to improper handling or maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18365979
Report Number1823260-2023-04136
Device Sequence Number18066690
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number04837738001
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer12/22/2023
Initial Report FDA Received Date12/20/2023
Supplement Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-