| Model Number |
407200 |
| Medical Device Problem Codes |
Device Contamination with Body Fluid (2317); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
12/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During an atrial fibrillation surgery, after the puncture was completed, the needle was taken out for irrigation and it was noted that a large mass of blood clots was washed out from the inside of the needle.The operation went smoothly and finally completed without adverse effects on the patient.The patient does not have a history of blood clots.The doctor alleges that the blood clot were caused by the puncture needle.
|
| |
|
Additional Manufacturer Narrative
|
|
One brk transseptal needle was received for evaluation.The stylet was not returned.No functional anomalies were noted when the needle was flushed and when an air aspiration test was performed using a slip tip syringe.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of patient blood clots within the needle remains unknown.
|
| |
|
Event or Problem Description
|
|
Fda request to update device model information and corrected reportable aware date.Follow up report required.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only**.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|
Event or Problem Description
|
|
This report includes updated device model information.
|
| |
|