| Catalog Number |
07.01702.015 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/04/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during an x-ray approximately 1 week postoperatively, the c7 left screw showed a crack, and an x-ray approximately 2 months post-operatively showed that the tulip head fractured on two screws.A revision surgery is scheduled to remove the construct and replace with competitive product.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 phone #: (b)(6).Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the weld on the lower housing fractured causing the screw to disassemble.X-rays were also provided and show that two of the implanted screws were fractured.The weld depth was also measured on the returned device and was found to be deeper than the minimum required weld depth of 0.010 inches.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Dhr review: per dhr review, the part was likely conforming when it left highridge control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report 3012447612-2023-00398.
|
| |
|
Event or Problem Description
|
|
It was reported that during an x-ray approximately 1 week postoperatively, the c7 left screw showed a crack, and an x-ray approximately 2 months post-operatively showed that the tulip head fractured on two screws.A revision surgery was performed to remove the construct and replace it with competitive product.This is report one of two for this event.
|
| |
|